What are the requirements for proper use and care of EBTs?

Title 49 C.F.R. ● ACTIVE Primary Source
Regulatory Text

49 C.F.R. § 40.233 — What are the requirements for proper use and care of EBTs?

§ 40.233 What are the requirements for proper use and care of EBTs? (a) As an EBT manufacturer, you must submit, for NHTSA approval, a quality assurance plan (QAP) for your EBT before ODAPC places the EBT on its Web page for “Approved Evidential Breath Measurement Devices.” (1) Your QAP must specify the methods used to perform external calibration checks on the EBT, the tolerances within which the EBT is regarded as being in proper calibration, and the intervals at which these checks must be performed. In designating these intervals, your QAP must take into account factors like frequency of use, environmental conditions ( e.g., temperature, humidity, altitude) and type of operation ( e.g., stationary or mobile). (2) Your QAP must also specify the inspection, maintenance, and calibration requirements and intervals for the EBT. (b) As the manufacturer, you must include, with each EBT, instructions for its use and care consistent with the QAP. (c) As the user of the EBT ( e.g., employer, service agent), you must do the following: (1) You must follow the manufacturer's instructions (see paragraph (b) of this section), including performance of external calibration checks at the intervals the instructions specify. (2) In conducting external calibration checks, you must use only calibration devices appearing on NHTSA's CPL for “Calibrating Units for Breath Alcohol Tests.” (3) If an EBT fails an external check of calibration, you must take the EBT out of service. You may not use the EBT again for DOT alcohol testing until it is repaired and passes an external calibration check. (4) You must maintain records of the inspection, maintenance, and calibration of EBTs as provided in § 40.333(a)(3). (5) You must ensure that inspection, maintenance, and calibration of the EBT are performed by its manufacturer or a maintenance representative certified either by the manufacturer or by a state health agency or other appropriate state agency. [65 FR 79526, Dec. 19, 2000, as amended at 82 FR 52246, Nov. 13, 2017]

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR49-SEC-1E9B3F
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
49 C.F.R. § 40.233
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
fe3477ad605bd538...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 49 C.F.R. § 40.233 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.
Navigate Corpus — Title 49 C.F.R.
◈ Provenance
boisestandard.org United States Law CFR Title 49 49 C.F.R. § 40.233