Capacity test requirements.

Title 42 C.F.R. ● ACTIVE Primary Source
Regulatory Text

42 C.F.R. § 84.304 — Capacity test requirements.

§ 84.304 Capacity test requirements. (a) NIOSH will conduct the capacity test on a total of 8 to 10 of the units submitted for approval, as follows: (1) Three units will be tested on a breathing and metabolic simulator in the condition in which they are received from the applicant; (2) Two units will be tested on a breathing and metabolic simulator after being subjected to the environmental treatments specified in § 84.307 of this subpart; (3) Two units will be tested on a breathing and metabolic simulator at the cold-temperature limit recommended by the manufacturer under § 84.302(h)(1), after the unit has been stored for a minimum of 24 hours at this limit; and (4) One unit, in the condition in which it is received from the applicant, will be tested by a human subject on a treadmill. (5) To approve a CCER for use in coal mines, two units will also be tested by a human subject under the specifications of §§ 84.99 and 84.100 that are applicable to man test 4. (b) The capacity test will begin upon the first inhalation from or exhalation into the unit. (c) Each unit will be tested at a constant work rate, depending on the capacity value specified by the manufacturer, according to the requirements specified in Table 2. All volumes are given at standard temperature (0 C) and pressure (760 mm Hg), dry, unless otherwise noted. (d) NIOSH will rate an approved CCER using the appropriate capacity rating, as specified in Table 2. Table 2—Capacity Test Requirements Capacity rating Capacity (L of O 2 ) VO 2 (L/min) VCO 2 (L/min) Ve (L/min) RF (Breaths/min) Cap 1 20 ≤L ≤59 2.50 2.50 55 22 Cap 2 60 ≤L ≤79 2.00 1.80 44 20 Cap 3 L ≥80 1.35 1.15 30 18 VO 2 = volume of oxygen consumed per minute; VCO 2 = volume of carbon dioxide produced per minute. Ve = ventilation rate in liters of air per minute; RF = respiratory frequency. (e) NIOSH will document the least value achieved by the seven units tested using the breathing and metabolic simulator. NIOSH will quantify this value of achieved capacity within an increment of 5 liters, rounding intermediate values to the nearest lower 5-liter increment.

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR42-SEC-12AA5F
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
42 C.F.R. § 84.304
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
778994c18c7fb57a...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 42 C.F.R. § 84.304 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.
Navigate Corpus — Title 42 C.F.R.
◈ Provenance
boisestandard.org United States Law CFR Title 42 42 C.F.R. § 84.304