42 C.F.R. § 84.303 — General testing conditions and requirements.

§ 84.303 General testing conditions and requirements. (a) NIOSH will conduct capacity and performance tests on the CCER using a breathing and metabolic simulator to provide quantitative evaluations and human subjects on a treadmill to provide qualitative evaluations. Information on the design and operation of the simulator is available from the NIOSH Web site at http://www.cdc.gov/niosh/npptl. Technical specifications can be obtained from NIOSH by contacting the National Personal Protective Technology Laboratory (NPPTL) by mail: P.O. Box 18070, 626 Cochrans Mill Road, Pittsburgh, PA 15236. Telephone: 412–386–4000 (this is not a toll-free number). Email: npptl@cdc.gov. (b) Capacity, performance, and wearability tests will continuously monitor the stressors listed in Table 1. The stressors and their respective acceptable ranges will be measured at the interface between the CCER and the mouth by instruments capable of breath-by-breath measurement. Stressor measurements will be evaluated as 1-minute averages. The operating averages of each stressor will be calculated upon the completion of each test as the average of the 1-minute measurements of the stressor recorded during the test. The level of any excursion for a stressor occurring during a test will be defined by the 1-minute average value(s) of the excursion(s). Table 1—Monitored Stressors and Their Acceptable Ranges Stressor Acceptable range operating average Acceptable range excursion Average inhaled CO 2 <1.5% ≤4%. Average inhaled O 2 >19.5% ≥15%. Peak Breathing Pressures ΔP ≤200 mm H 2 O −300 ≤ΔP ≤200 mm H 2 O. Wet-bulb temperature 1 <43 °C ≤50 °C. 1  Wet-bulb temperature is a measurement of the temperature of a wet surface. It represents the temperature of the inhaled breathing gas in the CCER user's trachea. (c) Capacity and performance tests will conclude when the stored breathing gas supply has been fully expended. (d) NIOSH will determine a CCER to have failed a capacity, performance, or wearability test if any of the following occurs: (1) A 1-minute average measurement of any stressor listed in Table 1 occurs outside the acceptable excursion range specified in Table 1; or an average stressor measurement calculated at the completion of a performance or capacity test exceeds the acceptable operating average range specified in Table 1; or (2) A human subject cannot complete the test for any reason related to the CCER, as determined by NIOSH. (e) Unless otherwise stated, tests required under this subpart will be conducted at the following ambient conditions: (1) Ambient temperatures of 23 °C ±3 °C; and (2) Atmospheric pressures of 735 mm Hg ±15 mm Hg.

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR42-SEC-9E50F6
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
42 C.F.R. § 84.303
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
34e9e6b8fd298ce2...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 42 C.F.R. § 84.303 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.
Navigate Corpus — Title 42 C.F.R.
◈ Provenance
boisestandard.org United States Law CFR Title 42 42 C.F.R. § 84.303