10 C.F.R. § 35.633 — Full calibration measurements on remote afterloader units.

§ 35.633 Full calibration measurements on remote afterloader units. (a) A licensee authorized to use a remote afterloader unit for medical use shall perform full calibration measurements on each unit— (1) Before the first medical use of the unit; (2) Before medical use under the following conditions: (i) Following replacement of the source or following reinstallation of the unit in a new location outside the facility; and (ii) Following any repair of the unit that includes removal of the source or major repair of the components associated with the source exposure assembly; and (3) At intervals not exceeding 1 quarter for high dose-rate, medium dose-rate, and pulsed dose-rate remote afterloader units with sources whose half-life exceeds 75 days; and (4) At intervals not exceeding 1 year for low dose-rate remote afterloader units. (b) To satisfy the requirement of paragraph (a) of this section, full calibration measurements must include, as applicable, determination of: (1) The output within ±5 percent; (2) Source positioning accuracy to within ±1 millimeter; (3) Source retraction with backup battery upon power failure; (4) Length of the source transfer tubes; (5) Timer accuracy and linearity over the typical range of use; (6) Length of the applicators; and (7) Function of the source transfer tubes, applicators, and transfer tube-applicator interfaces. (c) A licensee shall use the dosimetry system described in § 35.630(a) to measure the output. (d) A licensee shall make full calibration measurements required by paragraph (a) of this section in accordance with published protocols accepted by nationally recognized bodies. (e) In addition to the requirements for full calibrations for low dose-rate remote afterloader units in paragraph (b) of this section, a licensee shall perform an autoradiograph of the source(s) to verify inventory and source(s) arrangement at intervals not exceeding 1 quarter. (f) For low dose-rate remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made in accordance with paragraphs (a) through (e) of this section. (g) A licensee shall mathematically correct the outputs determined in paragraph (b)(1) of this section for physical decay at intervals consistent with 1 percent physical decay. (h) Full calibration measurements required by paragraph (a) of this section and physical decay corrections required by paragraph (g) of this section must be performed by the authorized medical physicist. (i) A licensee shall retain a record of each calibration in accordance with § 35.2632.

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR10-SEC-8CBBA2
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
10 C.F.R. § 35.633
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
aa5533b82a07445f...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 10 C.F.R. § 35.633 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.
Navigate Corpus — Title 10 C.F.R.
◈ Provenance
boisestandard.org United States Law CFR Title 10 10 C.F.R. § 35.633