10 C.F.R. § 35.632 — Full calibration measurements on teletherapy units.

§ 35.632 Full calibration measurements on teletherapy units. (a) A licensee authorized to use a teletherapy unit for medical use shall perform full calibration measurements on each teletherapy unit— (1) Before the first medical use of the unit; and (2) Before medical use under the following conditions: (i) Whenever spot-check measurements indicate that the output differs by more than 5 percent from the output obtained at the last full calibration corrected mathematically for radioactive decay; (ii) Following replacement of the source or following reinstallation of the teletherapy unit in a new location; (iii) Following any repair of the teletherapy unit that includes removal of the source or major repair of the components associated with the source exposure assembly; and (3) At intervals not exceeding 1 year. (b) To satisfy the requirement of paragraph (a) of this section, full calibration measurements must include determination of— (1) The output within ±3 percent for the range of field sizes and for the distance or range of distances used for medical use; (2) The coincidence of the radiation field and the field indicated by the light beam localizing device; (3) The uniformity of the radiation field and its dependence on the orientation of the useful beam; (4) Timer accuracy and linearity over the range of use; (5) On-off error; and (6) The accuracy of all distance measuring and localization devices in medical use. (c) A licensee shall use the dosimetry system described in § 35.630(a) to measure the output for one set of exposure conditions. The remaining radiation measurements required in paragraph (b)(1) of this section may be made using a dosimetry system that indicates relative dose rates. (d) A licensee shall make full calibration measurements required by paragraph (a) of this section in accordance with published protocols accepted by nationally recognized bodies. (e) A licensee shall mathematically correct the outputs determined in paragraph (b)(1) of this section for physical decay for intervals not exceeding 1 month for cobalt-60, 6 months for cesium-137, or at intervals consistent with 1 percent decay for all other nuclides. (f) Full calibration measurements required by paragraph (a) of this section and physical decay corrections required by paragraph (e) of this section must be performed by the authorized medical physicist. (g) A licensee shall retain a record of each calibration in accordance with § 35.2632.

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR10-SEC-447B23
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
10 C.F.R. § 35.632
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
8f4473c4ad1efec8...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 10 C.F.R. § 35.632 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.
Navigate Corpus — Title 10 C.F.R.
◈ Provenance
boisestandard.org United States Law CFR Title 10 10 C.F.R. § 35.632