42 C.F.R. · Public Health
42 C.F.R. § 84.204

Exhalation valve leakage test; minimum requirements.

Title 42 C.F.R. ● ACTIVE Primary Source
Regulatory Text

42 C.F.R. § 84.204 — Exhalation valve leakage test; minimum requirements.

§ 84.204 Exhalation valve leakage test; minimum requirements. (a) Dry exhalation valves and valve seats will be subjected to a suction of 25 mm. water-column height while in a normal operating position. (b) Leakage between the valve and valve seat shall not exceed 30 milliliters per minute.

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR42-SEC-83DC9C
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
42 C.F.R. § 84.204
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
d0127a104151dec1...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 42 C.F.R. § 84.204 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.