30 C.F.R. § 27.34 — Test for intrinsic safety.
§ 27.34 Test for intrinsic safety. Assemblies, subassemblies, or components that are designed for intrinsic safety shall be tested by introducing into the circuit(s) thereof a circuit-interrupting device which produces an electric spark from the current in the circuit. The circuit-interrupting device shall be placed in a gallery containing various flammable natural gas-air mixtures. To meet the requirements of this test, the spark shall not ignite the flammable mixture. For this test the circuit-interrupting device shall be operated not less than 100 times at 125 percent of the normal operating voltage of the particular circuit.
Source: ecfr.gov · govinfo.gov — public domain
Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record
Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR30-SEC-CF6ECC
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
30 C.F.R. § 27.34
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
4b96d251efccfbb2...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 30 C.F.R. § 27.34 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.
Navigate Corpus — Title 30 C.F.R.
Title 30 — Mineral Resources
30 C.F.R. § 1.130 C.F.R. § 1.230 C.F.R. § 1.330 C.F.R. § 3.130 C.F.R. § 5.1030 C.F.R. § 5.3030 C.F.R. § 5.4030 C.F.R. § 5.5030 C.F.R. § 6.130 C.F.R. § 6.230 C.F.R. § 6.1030 C.F.R. § 6.2030 C.F.R. § 6.3030 C.F.R. § 7.130 C.F.R. § 7.230 C.F.R. § 7.330 C.F.R. § 7.430 C.F.R. § 7.530 C.F.R. § 7.630 C.F.R. § 7.730 C.F.R. § 7.830 C.F.R. § 7.930 C.F.R. § 7.1030 C.F.R. § 7.2130 C.F.R. § 7.2230 C.F.R. § 7.2330 C.F.R. § 7.2430 C.F.R. § 7.2530 C.F.R. § 7.2630 C.F.R. § 7.27
30 C.F.R. — public domain
← All Legislation