21 C.F.R. § 874.5900 — External upper esophageal sphincter compression device.

§ 874.5900 External upper esophageal sphincter compression device. (a) Identification. An external upper esophageal sphincter compression device is a prescription device used to apply external pressure on the cricoid cartilage for the purpose of reducing the symptoms of laryngopharyngeal reflux disease. (b) Classification. Class II (special controls). The special controls for this device are: (1) The patient contacting components must be demonstrated to be biocompatible. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated: (i) Mechanical integrity testing ( e.g., tensile strength testing, fatigue testing) and (ii) Shelf life testing. (3) The technical specifications must include pressure measurement accuracy to characterize device performance. (4) Clinical performance testing must document any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use. (5) Labeling must include the following: (i) Appropriate warnings and precautions, (ii) A detailed summary of the clinical testing pertinent to use of the device including a detailed summary of the device-related complications or adverse events, (iii) Detailed instructions on how to fit the device to the patient, and (iv) Instructions for reprocessing of any reusable components. (6) Patient labeling must be provided and must include: (i) Relevant warnings, precautions, and adverse effects/complications, (ii) Information on how to correctly wear the device, (iii) The potential risks and benefits associated with the use of the device, (iv) Alternative treatments, and (v) Reprocessing instructions. [80 FR 46194, Aug. 4, 2015]

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR21-SEC-FF97D0
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
21 C.F.R. § 874.5900
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
2c5209760f0aa521...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 21 C.F.R. § 874.5900 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.