21 C.F.R. · Food and Drugs
21 C.F.R. § 870.4150

Extracorporeal system for carbon dioxide removal.

Title 21 C.F.R. ● ACTIVE Primary Source
Regulatory Text

21 C.F.R. § 870.4150 — Extracorporeal system for carbon dioxide removal.

§ 870.4150 Extracorporeal system for carbon dioxide removal. (a) Identification. An extracorporeal system for carbon dioxide removal is a system of devices and accessories that provides assisted extracorporeal carbon dioxide removal from the patient's blood in patients with acute respiratory failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, a gas exchanger, blood pump, cannulae, tubing, filters, and other accessories ( e.g., monitors, detectors, sensors, connectors). (b) Classification. Class II (special controls). The special controls for this device are: (1) In vivo evaluation, which may include animal testing and clinical data, of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the in vivo evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness. (2) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible. (3) Non-clinical performance testing of the devices and accessories in the circuit must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Mechanical integrity; (ii) Durability; and (iii) Reliability. (4) All patient contacting components of the device must be demonstrated to be biocompatible. (5) Performance testing must demonstrate the electrical safety and electromagnetic compatibility (EMC) of any electrical components. (6) Software validation, verification, and hazard analysis must be performed. (7) Performance testing must demonstrate the sterility of all patient-contacting components. (8) Performance testing must support the shelf life of the device by demonstrating continued sterility and device functionality over the identified shelf life. (9) Labeling must include the following: (i) A detailed summary of the non-clinical and in vivo evaluations pertinent to use of the device and accessories in the circuit; (ii) Adequate instructions with respect to circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure; and (iii) A shelf life. [87 FR 80039, Dec. 29, 2022]

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR21-SEC-C13F4B
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
21 C.F.R. § 870.4150
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
37d1b6a3110e84c6...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 21 C.F.R. § 870.4150 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.
Navigate Corpus — Title 21 C.F.R.
◈ Provenance
boisestandard.org United States Law CFR Title 21 21 C.F.R. § 870.4150