21 C.F.R. · Food and Drugs
21 C.F.R. § 868.5273

Positive airway pressure delivery system.

Title 21 C.F.R. ● ACTIVE Primary Source
Regulatory Text

21 C.F.R. § 868.5273 — Positive airway pressure delivery system.

§ 868.5273 Positive airway pressure delivery system. (a) Identification. A positive airway pressure delivery system is a prescription noninvasive ventilatory device that delivers expiratory positive airway pressure for patients suffering from obstructive sleep apnea. The system also provides positive airway pressure during incipient apnea. The system may include a dedicated flow generator and a patient interface. (b) Classification. Class II (special controls). The special controls for this device are: (1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following: (i) Waveform testing must simulate breathing conditions and evaluate pressure and airflow response over a range and combination of high and low breath rates and tidal volumes. (ii) Use life testing must demonstrate adequate device performance over the labeled use life of the device. (iii) Device integrity testing must demonstrate that the device can withstand typical forces expected during use. (iv) Carbon dioxide rebreathing testing must be performed. (v) System flow rate, maximum expiratory pressure, inhalation pressure, and intra-mask static pressure testing must be performed. (vi) Air bolus testing must demonstrate that the device can withstand worst-case scenario air pressures. (vii) Maximum limited pressure testing of the flow generator in single fault condition must be performed. (viii) Maximum output temperature testing of delivered gas, if humidified, must be performed. (3) Performance data must validate reprocessing instructions for any reusable components of the device. (4) Performance data must demonstrate the electrical, thermal, and mechanical safety and the electromagnetic compatibility of the device. (5) Software verification, validation, and hazard analysis must be performed. (6) Labeling must include the following: (i) Therapy pressure range; (ii) Use life and replacement schedule for all components; (iii) Cleaning instructions; and (iv) Instructions for assembly and connection of device components. [83 FR 52966, Oct. 19, 2018]

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR21-SEC-4E7B6A
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
21 C.F.R. § 868.5273
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
bb9dc3a8febce624...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 21 C.F.R. § 868.5273 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.