21 C.F.R. · Food and Drugs
21 C.F.R. § 812.43

Selecting investigators and monitors.

Title 21 C.F.R. ● ACTIVE Primary Source
Regulatory Text

21 C.F.R. § 812.43 — Selecting investigators and monitors.

§ 812.43 Selecting investigators and monitors. (a) Selecting investigators. A sponsor shall select investigators qualified by training and experience to investigate the device. (b) Control of device. A sponsor shall ship investigational devices only to qualified investigators participating in the investigation. (c) Obtaining agreements. A sponsor shall obtain from each participating investigator a signed agreement that includes: (1) The investigator's curriculum vitae. (2) Where applicable, a statement of the investigator's relevant experience, including the dates, location, extent, and type of experience. (3) If the investigator was involved in an investigation or other research that was terminated, an explanation of the circumstances that led to termination. (4) A statement of the investigator's commitment to: (i) Conduct the investigation in accordance with the agreement, the investigational plan, this part and other applicable FDA regulations, and conditions of approval imposed by the reviewing IRB or FDA; (ii) Supervise all testing of the device involving human subjects; and (iii) Ensure that the requirements for obtaining informed consent are met. (5) Sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement as required under part 54 of this chapter. The sponsor shall obtain a commitment from the clinical investigator to promptly update this information if any relevant changes occur during the course of the investigation and for 1 year following completion of the study. This information shall not be submitted in an investigational device exemption application, but shall be submitted in any marketing application involving the device. (d) Selecting monitors. A sponsor shall select monitors qualified by training and experience to monitor the investigational study in accordance with this part and other applicable FDA regulations. [45 FR 3751, Jan. 18, 1980, as amended at 63 FR 5253, Feb. 2, 1998]

Source: ecfr.gov · govinfo.gov — public domain Official Source ↗
Root-LD Entity Data
◈ Machine-Readable Provenance Record Root-LD v1.0 · boisestandard.org
Federation ID
BS-CFR21-SEC-DCD8F9
Entity Class
REGULATION / FEDERAL-CFR-SECTION
Domain Signature
boisestandard.org
Citation
21 C.F.R. § 812.43
Corpus
CFR — Code of Federal Regulations
Status
✓ ACTIVE
Source
PRIMARY-SOURCE
Source Verified
✓ TRUE
Content Hash
b17cb17f7b766cfe...
Semantic Edges
Pending — corpus passes queued
The regulatory text of 21 C.F.R. § 812.43 is reproduced from the official Code of Federal Regulations as published by the Office of the Federal Register and the National Archives and Records Administration.